Introduction

What you need

Step 1

              Opening the Column in the Hamilton Galileo               
  • Use a grounding cable and wrist band connected to the GALILEO or other work surface when you are working on the GALILEO.

Use a grounding cable and wrist band connected to the GALILEO or other work surface when you are working on the GALILEO.

1024

Step 2

  • Remove from GALILEO: External mains power supply, Air and oxygen supplies, Patient breathing circuit (for convenience only), Flow Sensor tubes (for convenience only).

Remove from GALILEO: External mains power supply, Air and oxygen supplies, Patient breathing circuit (for convenience only), Flow Sensor tubes (for convenience only).

Step 3

  • Loosen the rear panel of the column by removing the six hexagonal-drive (Allen) screws that hold it in place.

Loosen the rear panel of the column by removing the six hexagonal-drive (Allen) screws that hold it in place.

Step 4

  • Without removing any electrical connections, open the left side of the column as wide as practical.
  • This exposes: GALILEO’s internal power supply, The battery fuse, The diagnostic LEDs on the power supply’s main circuit board, and the back up batteries.

Without removing any electrical connections, open the left side of the column as wide as practical.

This exposes: GALILEO’s internal power supply, The battery fuse, The diagnostic LEDs on the power supply’s main circuit board, and the back up batteries.

Step 5

  • Make sure no electrical connections are touching the column.
  • Make sure all green/brown grounding (earthing) cables are securely in place.

Make sure no electrical connections are touching the column.

Make sure all green/brown grounding (earthing) cables are securely in place.

To reassemble your device, follow these instructions in reverse order.

Cancel: I did not complete this guide.

Author

                                      with 1 other contributor 

                    Ian Van Sciver                     

Member since: 04/09/2020

332 Reputation

                                      2 Guides authored                  



                       Badges:
                       5







                                                        +2 more badges                           

Team

                       Cal Poly, Team S7-G26, Paton Spring 2020                        

                                                  Member of Cal Poly, Team S7-G26, Paton Spring 2020 



                    CPSU-PATON-S20S7G26                     


                                            3 Members                     


                                            4 Guides authored