Introduction
The humidifier feature should be checked every few months to ensure that it is working properly. The following guide details how to take all of these parts apart and replace them if they are malfunctioning.
For more information on replacing the humidifier, please refer to page 7-2 in the service manual.
What you need
Step 1
Humidifier
- Remove the Water Chamber Assembly.
- Gently squeeze the latch on the Flip Lid Assembly to release it.
- Lift the Flip Lid Assembly.
- Pull the Humidifier Tank Assembly out of the Humidifier.
Remove the Water Chamber Assembly.
Gently squeeze the latch on the Flip Lid Assembly to release it.
Lift the Flip Lid Assembly.
Pull the Humidifier Tank Assembly out of the Humidifier.
1024
Step 2
- Remove the Tank Top Seal from the Flip Lid Assembly.
- If this is the source of the error, put in a new one.
Remove the Tank Top Seal from the Flip Lid Assembly.
If this is the source of the error, put in a new one.
Step 3
- Remove the Dry Box Seal.
- If this is the source of error, place this part with a new one.
- When replacing, ensure that the narrow end of the seal is inserted into the Dry Box.
Remove the Dry Box Seal.
If this is the source of error, place this part with a new one.
When replacing, ensure that the narrow end of the seal is inserted into the Dry Box.
Step 4
- Remove the Dry Box Assembly.
- Insert a flat blade screwdriver into the hole located on the back of the Humidifier and lightly press inward to release the Dry Box Assembly.
- Do not press firmly on the screwdriver as damage may occur.
- Remove the Inlet Seal from the Dry Box Assembly
Remove the Dry Box Assembly.
Insert a flat blade screwdriver into the hole located on the back of the Humidifier and lightly press inward to release the Dry Box Assembly.
Do not press firmly on the screwdriver as damage may occur.
Remove the Inlet Seal from the Dry Box Assembly
Step 5
- Remove the Flip Lid Assembly.
- Use a screwdriver to push in on the latches that secure the Flip Lid Assembly to the Humidifier Top Housing.
- Bend the Flip Lid Assembly completely backwards until it is removed.
- For Heated Tube devices, remove the wire harness from the base.
Remove the Flip Lid Assembly.
Use a screwdriver to push in on the latches that secure the Flip Lid Assembly to the Humidifier Top Housing.
Bend the Flip Lid Assembly completely backwards until it is removed.
For Heated Tube devices, remove the wire harness from the base.
Step 6
- Use a T8 Torx screwdriver to remove the four #4 x 1/2" screws that secure the Top Housing to the Humidifier Bottom Housing
Use a T8 Torx screwdriver to remove the four #4 x 1/2" screws that secure the Top Housing to the Humidifier Bottom Housing
Step 7
- Slide the Outside Cover out of the Bottom Housing.
Slide the Outside Cover out of the Bottom Housing.
Step 8
- Remove the Left Side Panel.
- Use a T15 Torx screwdriver to remove the four #6 x 1/4" screws that secure the Bottom Housing to the Lower Base Assembly.
Remove the Left Side Panel.
Use a T15 Torx screwdriver to remove the four #6 x 1/4" screws that secure the Bottom Housing to the Lower Base Assembly.
Step 9
- Remove the Heater Plate Assembly.
- The Heater Plate Assembly is secured by two small tabs.
- Once replaced, secure the Bottom Housing to the Lower Base using the four #6 x 1/4" screws.
- Be sure the Heater Plate Spring is properly seated under the Heater Plate.
Remove the Heater Plate Assembly.
The Heater Plate Assembly is secured by two small tabs.
Once replaced, secure the Bottom Housing to the Lower Base using the four #6 x 1/4" screws.
Be sure the Heater Plate Spring is properly seated under the Heater Plate.
To reassemble your device, follow these instructions in reverse order.
Cancel: I did not complete this guide.
Attached Documents
- Respironics_BiPAP-ST_Ventilator_-_Service_manual
- PDF - 85.50 kb
- View
Respironics_BiPAP-ST_Ventilator_-_Service_manual
PDF - 85.50 kb
Author
with 2 other contributors
Courtney Huitt
Member since: 04/07/2020
169 Reputation
1 Guide authored
Badges:
2
Team
Cal Poly, Team S7-G3, Paton Spring 2020
Member of Cal Poly, Team S7-G3, Paton Spring 2020
CPSU-PATON-S20S7G3
3 Members
3 Guides authored