  • Philips Emerson CoughAssist CA Series Disassembly
  • 6 steps
  • Valve Actuator Motor
  • 4 steps

Flags

1

Needs More Images

A few more images would make this guide’s procedures crystal clear.

  • BackPhilips Emerson CoughAssist CA Series

  • Full Screen

  • Options

  • History

  • Save to Favorites

  • Download PDF

  • Edit

  • Translate

  • Get Shareable Link

  • Embed This Guide

  • Notify Me of Changes

  • Stop Notifications

Introduction

What you need

Step 1

              Philips Emerson CoughAssist CA Series Disassembly               
  • First, remove the two 16mm pan head screws (1) directly under the front panel of the ventilator. (Page 3).

First, remove the two 16mm pan head screws (1) directly under the front panel of the ventilator. (Page 3).

1024

Step 2

  • Next, remove the two 16mm flat head screws (3) from the back panel of the ventilator. (Page 3).

Next, remove the two 16mm flat head screws (3) from the back panel of the ventilator. (Page 3).

Step 3

  • Lift off the top cover (2) of the device. (Page 3).

Lift off the top cover (2) of the device. (Page 3).

Step 4

  • Now remove the “Inhale flow” control knob (4) and “Inhale pressure” knob (4), by loosening the setscrew on each before pulling off. If you have a CM model, carefully remove the Manual Control Lever (3) by prying it off with the screwdriver. (Page 3).

Now remove the “Inhale flow” control knob (4) and “Inhale pressure” knob (4), by loosening the setscrew on each before pulling off. If you have a CM model, carefully remove the Manual Control Lever (3) by prying it off with the screwdriver. (Page 3).

Step 5

  • Unscrew the three 12.7mm plastite screws ( 2) holding the panel assembly to the lower housing. A long Phillips screwdriver is very helpful for access to the three screws. (Page 3)

Unscrew the three 12.7mm plastite screws ( 2) holding the panel assembly to the lower housing. A long Phillips screwdriver is very helpful for access to the three screws. (Page 3)

Step 6

  • The front control panel (5) should now be loose. Carefully slide it forwards to remove. (Page 4).

The front control panel (5) should now be loose. Carefully slide it forwards to remove. (Page 4).

Step 7

              Valve Actuator Motor               
  • Position the ventilator so the pressure gauge is facing right to your right.
  • Follow the Preliminary Disassembly Guide to remove the top cover and the front panel.

Position the ventilator so the pressure gauge is facing right to your right.

Follow the Preliminary Disassembly Guide to remove the top cover and the front panel.

Step 8

  • Remove the two 12.7mm plastite screws (10) as shown, and remove the Power Filter (9). (Page 7)
  • Remove all screws holding the Bracket (6) to the Valve Assembly (12). (Page 7)
  • Remove the two 12.7mm plastite screws above and the one screw beneath the Actuator Motor (9). (Page 7)
  • Take out the Actuator Motor (9) while it is still attached to the Bracket (6). (Page 7)

Remove the two 12.7mm plastite screws (10) as shown, and remove the Power Filter (9). (Page 7)

Remove all screws holding the Bracket (6) to the Valve Assembly (12). (Page 7)

Remove the two 12.7mm plastite screws above and the one screw beneath the Actuator Motor (9). (Page 7)

Take out the Actuator Motor (9) while it is still attached to the Bracket (6). (Page 7)

Step 9

  • If you notice other parts are damaged or faulty, replace them. These parts include the torsion spring (1), cable chain (3), larger sprocket (4), or the spacer (2). (Page 7).
  • The small sprocket on the motor is not designed to be removed from the shaft, so make sure you are replacing the large sprocket.

If you notice other parts are damaged or faulty, replace them. These parts include the torsion spring (1), cable chain (3), larger sprocket (4), or the spacer (2). (Page 7).

The small sprocket on the motor is not designed to be removed from the shaft, so make sure you are replacing the large sprocket.

Step 10

  • Disconnect the two motor terminals from the motor capacitor (7). (Page 7).
  • Remove the 4 6.4mm binding head screws holding the motor to the bracket. (Page 7).
  • Remove the motor. (Page 7)

Disconnect the two motor terminals from the motor capacitor (7). (Page 7).

Remove the 4 6.4mm binding head screws holding the motor to the bracket. (Page 7).

Remove the motor. (Page 7)

To reassemble your device, follow these instructions in reverse order. After reassembly, check that the cable chain(3) is tensioned properly. This means that the two sides of the chain can be pinched together almost half way, but not able to touch. If you need more detail or looking for more information on this repair, please refer to the Service Guide attached, page 7.

Cancel: I did not complete this guide.

Attached Documents

  • CoughAssist-CA-3000-CA-3200-CM.pdf
  • PDF - 481.56 kb
  • View

CoughAssist-CA-3000-CA-3200-CM.pdf

PDF - 481.56 kb

Author

                                      with 2 other contributors 

                    Zachary Stednitz                     

Member since: 04/07/2020

292 Reputation

                                      2 Guides authored                  



                       Badges:
                       5





                                                        +2 more badges                           

Team

                       Cal Poly, Team S11-G18, Paton Spring 2020                        

                                                  Member of Cal Poly, Team S11-G18, Paton Spring 2020 



                    CPSU-PATON-S20S11G18                     


                                            3 Members                     


                                            3 Guides authored