Introduction

Step 1

              Oxygen Flow Sensor/Valve               
  • If the ventilator is not already in diagnostic mode, simultaneously press the ALARM RESET and 100% O2 keys on the front panel for approximately five seconds while turning ventilator power on, then touch OK to enter diagnostic mode. (Service Manual 8-19)

If the ventilator is not already in diagnostic mode, simultaneously press the ALARM RESET and 100% O2 keys on the front panel for approximately five seconds while turning ventilator power on, then touch OK to enter diagnostic mode. (Service Manual 8-19)

1024

Step 2

  • Reconnect the oxygen source to the ventilator. (Service Manual 8-19)

Reconnect the oxygen source to the ventilator. (Service Manual 8-19)

Step 3

  • Touch Oxygen on the Hardware screen and set the flow to 1 LPM. Verify that the O2Position display reads from 185 to 525 steps. (Service Manual 8-19)

Touch Oxygen on the Hardware screen and set the flow to 1 LPM. Verify that the O2Position display reads from 185 to 525 steps. (Service Manual 8-19)

Step 4

  • Touch Oxygen and set the flow to 180 LPM. Check that the O2 Position display corresponds to the set altitude: (Service Manual 8-19)

Touch Oxygen and set the flow to 180 LPM. Check that the O2 Position display corresponds to the set altitude: (Service Manual 8-19)

Step 5

  • Connect a patient circuit and analyzer to the ventilator. (Service Manual 8-19)

Connect a patient circuit and analyzer to the ventilator. (Service Manual 8-19)

Step 6

  • Set the analyzer’s function to read LPM mode and gas select to O2 (set a high flow range capable of measuring 180 LPM oxygen). (Service Manual 8-19)

Set the analyzer’s function to read LPM mode and gas select to O2 (set a high flow range capable of measuring 180 LPM oxygen). (Service Manual 8-19)

Step 7

  • Check that set oxygen flow corresponds to measured flow: (Service Manual 8-19/20)

Check that set oxygen flow corresponds to measured flow: (Service Manual 8-19/20)

Step 8

  • The oxygen flow accuracy test is complete. (Service Manual 8-20)

The oxygen flow accuracy test is complete. (Service Manual 8-20)

To reassemble your device, follow these instructions in reverse order.

Cancel: I did not complete this guide.

Author

                                      with 1 other contributor 

                    Ethan Norman                     

Member since: 05/04/2020

157 Reputation

                                      1 Guide authored                  



                       Badges:
                       2

Team

                       Cal Poly, Team S1-G4, Zimmerman Spring 2020                        

                                                  Member of Cal Poly, Team S1-G4, Zimmerman Spring 2020 



                    CPSU-ZIMMERMAN-S20S1G4                     


                                            5 Members                     


                                            5 Guides authored