Introduction
Step 1
Oxygen Flow Sensor/Valve
- If the ventilator is not already in diagnostic mode, simultaneously press the ALARM RESET and 100% O2 keys on the front panel for approximately five seconds while turning ventilator power on, then touch OK to enter diagnostic mode. (Service Manual 8-19)
If the ventilator is not already in diagnostic mode, simultaneously press the ALARM RESET and 100% O2 keys on the front panel for approximately five seconds while turning ventilator power on, then touch OK to enter diagnostic mode. (Service Manual 8-19)
1024
Step 2
- Reconnect the oxygen source to the ventilator. (Service Manual 8-19)
Reconnect the oxygen source to the ventilator. (Service Manual 8-19)
Step 3
- Touch Oxygen on the Hardware screen and set the flow to 1 LPM. Verify that the O2Position display reads from 185 to 525 steps. (Service Manual 8-19)
Touch Oxygen on the Hardware screen and set the flow to 1 LPM. Verify that the O2Position display reads from 185 to 525 steps. (Service Manual 8-19)
Step 4
- Touch Oxygen and set the flow to 180 LPM. Check that the O2 Position display corresponds to the set altitude: (Service Manual 8-19)
Touch Oxygen and set the flow to 180 LPM. Check that the O2 Position display corresponds to the set altitude: (Service Manual 8-19)
Step 5
- Connect a patient circuit and analyzer to the ventilator. (Service Manual 8-19)
Connect a patient circuit and analyzer to the ventilator. (Service Manual 8-19)
Step 6
- Set the analyzer’s function to read LPM mode and gas select to O2 (set a high flow range capable of measuring 180 LPM oxygen). (Service Manual 8-19)
Set the analyzer’s function to read LPM mode and gas select to O2 (set a high flow range capable of measuring 180 LPM oxygen). (Service Manual 8-19)
Step 7
- Check that set oxygen flow corresponds to measured flow: (Service Manual 8-19/20)
Check that set oxygen flow corresponds to measured flow: (Service Manual 8-19/20)
Step 8
- The oxygen flow accuracy test is complete. (Service Manual 8-20)
The oxygen flow accuracy test is complete. (Service Manual 8-20)
To reassemble your device, follow these instructions in reverse order.
Cancel: I did not complete this guide.
Author
with 1 other contributor
Ethan Norman
Member since: 05/04/2020
157 Reputation
1 Guide authored
Badges:
2
Team
Cal Poly, Team S1-G4, Zimmerman Spring 2020
Member of Cal Poly, Team S1-G4, Zimmerman Spring 2020
CPSU-ZIMMERMAN-S20S1G4
5 Members
5 Guides authored