Introduction

Information in this guide was sourced from Chapter 10.1.3 on page 10-5 of the Respironics V200/Esprit Ventilator Service Manual. Further information on general filter replacements for this Ventilator can be found in Chapter 9.1 on pages 9-3 and 9-4 of the Service Manual.

What you need

Step 1

              Oxygen Inlet Filter               
  • Turn off the device.
  • On/Off switch located on the Front Panel as seen in Figure 8-2 (from the Operator Manual).

Turn off the device.

On/Off switch located on the Front Panel as seen in Figure 8-2 (from the Operator Manual).

1024

Step 2

  • Disconnect from AC power.

Disconnect from AC power.

Step 3

  • Disconnect the oxygen hose.
  • Oxygen hose is connected to the Inlet Connector (section 14 of Figure 5-1, diagram from Operator Manual).

Disconnect the oxygen hose.

Oxygen hose is connected to the Inlet Connector (section 14 of Figure 5-1, diagram from Operator Manual).

Step 4

  • Unscrew the water trap bowl (page 10-5 in Service Manual).
  • Located on the back panel of the ventilator (section 15 of Figure 5-1 from Operator Manual).

Unscrew the water trap bowl (page 10-5 in Service Manual).

Located on the back panel of the ventilator (section 15 of Figure 5-1 from Operator Manual).

Step 5

  • Separate the filter element from the black plastic retainer (page 10-5 of Service Manual).

Separate the filter element from the black plastic retainer (page 10-5 of Service Manual).

Step 6

  • Attach new filter element to the black plastic retainer (page 10-5 of Service Manual).

Attach new filter element to the black plastic retainer (page 10-5 of Service Manual).

Step 7

  • Screw on water trap bowl (page 10-5 of Service Manual).

Screw on water trap bowl (page 10-5 of Service Manual).

Step 8

  • Reconnect the oxygen hose.
  • According to page 2-2 of the Operators Manual, care in the routing of the oxygen inlet hose should be exercised to ensure it is not exposed to mechanisms that could cause damage by cutting or heating/melting.

Reconnect the oxygen hose.

According to page 2-2 of the Operators Manual, care in the routing of the oxygen inlet hose should be exercised to ensure it is not exposed to mechanisms that could cause damage by cutting or heating/melting.

  • Reconnect to AC Power.To reassemble your device, follow these instructions in reverse order.

Cancel: I did not complete this guide.

Author

                                      with 2 other contributors 

                    Grace Gius                     

Member since: 05/04/2020

505 Reputation

                                      3 Guides authored                  



                       Badges:
                       6







                                                        +3 more badges                           

Team

                       Cal Poly, Team S1-G4, Zimmerman Spring 2020                        

                                                  Member of Cal Poly, Team S1-G4, Zimmerman Spring 2020 



                    CPSU-ZIMMERMAN-S20S1G4                     


                                            5 Members                     


                                            5 Guides authored