- chapter 7.14 “Component Removal/Installation - Speakers” of service manualchapter 7.10 “Component Removal/Installation - Front Panel” of service manualchapter 7.9 “Component Removal/Installation - Separating the UI from the Base” of service manualchapter 7.8“Component Removal/Installation - Motor Controller (MC) PCBA” of service manualchapter 7.4 “Component Removal/Installation - Top Cover” of service manual
What you need
Step 1
Alarm Speaker
- step 1 Remove MC PCBA
- Remove top cover (top cover guide here)
- Remove the screws and wing nuts that attach the MC PCBA to the PCBA brackets.
- Loosen the captive screws that attach the capacitor bracket to the base.
- Disconnect cables: • Blower cable • Fan cable • DA-MC ribbon cable
- Gently pull MC PCBA up from its connection on the CPU PCBA, then disconnect speaker #1 cable.
- Remove PCBA brackets. To reinstall, follow the steps above in reverse order
step 1 Remove MC PCBA
Remove top cover (top cover guide here)
Remove the screws and wing nuts that attach the MC PCBA to the PCBA brackets.
Loosen the captive screws that attach the capacitor bracket to the base.
Disconnect cables: • Blower cable • Fan cable • DA-MC ribbon cable
Gently pull MC PCBA up from its connection on the CPU PCBA, then disconnect speaker #1 cable.
Remove PCBA brackets. To reinstall, follow the steps above in reverse order
1024
Step 2
- step 2 separate the UI from the base
- Remove MC PCBA (step 1)
- Disconnect cables: • LCD cable from PM PCBA. • LCD cable ground wires from the ground terminal. • LCD tray ground wires from the ground terminal. • UI cable (ribbon cable) from the PM PCBA.
- Loosen the captive screw that attaches the UI retainer to the UI.
- Slide retainer up, then remove UI from base.
step 2 separate the UI from the base
Remove MC PCBA (step 1)
Disconnect cables: • LCD cable from PM PCBA. • LCD cable ground wires from the ground terminal. • LCD tray ground wires from the ground terminal. • UI cable (ribbon cable) from the PM PCBA.
Loosen the captive screw that attaches the UI retainer to the UI.
Slide retainer up, then remove UI from base.
Step 3
- step 3 remove the front panel
- Separate UI from base (step 2)
- Remove hose clamp and boot that attach blower to the gas outlet.
- Remove the tubing from the barbed fittings inside the front panel.
- Disconnect gas outlet and proximal port ground wires from the base.
- Remove the screws that attach the front panel to the base and side walls.
- NOTE: When reinstalling tubing, ensure that: • Longer (9-in.) tube is connected between manifold barb “M” (machine) to barbed gas outlet fitting. • Shorter (8-in.) tube is connected between manifold barb “P” (proximal) and barbed proximal pressure port.
step 3 remove the front panel
Separate UI from base (step 2)
Remove hose clamp and boot that attach blower to the gas outlet.
Remove the tubing from the barbed fittings inside the front panel.
Disconnect gas outlet and proximal port ground wires from the base.
Remove the screws that attach the front panel to the base and side walls.
NOTE: When reinstalling tubing, ensure that: • Longer (9-in.) tube is connected between manifold barb “M” (machine) to barbed gas outlet fitting. • Shorter (8-in.) tube is connected between manifold barb “P” (proximal) and barbed proximal pressure port.
Step 4
- step 4 remove the alarm speakers
- Remove front panel (step 3)
- Disconnect speaker #2 from the power management (PM) PCBA. (Speaker #1 is disconnected when the MC PCBA is disconnected.)
- Remove the screw and washer that attaches each speaker to the base.
- Insert new speaker in place of old one
- Follow step 4 in reverse order to install speaker.
step 4 remove the alarm speakers
Remove front panel (step 3)
Disconnect speaker #2 from the power management (PM) PCBA. (Speaker #1 is disconnected when the MC PCBA is disconnected.)
Remove the screw and washer that attaches each speaker to the base.
Insert new speaker in place of old one
Follow step 4 in reverse order to install speaker.
Step 5
- step 5 reinstalling
- Follow step 3 in reverse order to reinstall the front panel.
- Follow step 2 in reverse order to reinstall the UI to the base.
- Follow step 1 in reverse order to reinstall the MC PCBA.
- Follow the Top Cover Guide in reverse order to reinstall.
step 5 reinstalling
Follow step 3 in reverse order to reinstall the front panel.
Follow step 2 in reverse order to reinstall the UI to the base.
Follow step 1 in reverse order to reinstall the MC PCBA.
Follow the Top Cover Guide in reverse order to reinstall.
To reassemble your device, follow these instructions in reverse order.
Cancel: I did not complete this guide.
Author
with 1 other contributor
Weihan Du
Member since: 04/11/2020
241 Reputation
2 Guides authored
Badges:
4
+1 more badge
Team
Cal Poly, Team S7-G10, Paton Spring 2020
Member of Cal Poly, Team S7-G10, Paton Spring 2020
CPSU-PATON-S20S7G10
2 Members
9 Guides authored