Difficulty

Easy

Steps

6

Time Required

                          25 minutes            

Sections

3

  • How to power down a Philips Respironics V60
  • 2 steps
  • How to Disconnect a Philips Respironics V60
  • 2 steps
  • Top Cover
  • 2 steps

Flags

1

Member-Contributed Guide

An awesome member of our community made this guide. It is not managed by iFixit staff.

  • BackPhilips Respironics V60

  • Full Screen

  • Options

  • History

  • Save to Favorites

  • Download PDF

  • Edit

  • Translate

  • Get Shareable Link

  • Embed This Guide

  • Notify Me of Changes

  • Stop Notifications

Introduction

The top cover just simply serves as a cover for the inside of the ventilator, making sure that the parts of the ventilator within are protected. Damage to this could also damage the inside of the ventilator so it would be good to replace/repair it as quickly as possible.

What you need

Step 1

              shut down the ventilator               
  • Press and release the ON/Shutdown key, opening the Shutdown window in the screen.
  • Select Ventilator Shutdown, which will shut down the ventilator.
  • Improper shutdown may cause a Power has been restored message the next time the ventilator is turned on.

Press and release the ON/Shutdown key, opening the Shutdown window in the screen.

Select Ventilator Shutdown, which will shut down the ventilator.

Improper shutdown may cause a Power has been restored message the next time the ventilator is turned on.

1024

Step 2

              disconnect from power               
  • Disconnect the ventilator from the AC power.

Disconnect the ventilator from the AC power.

Step 3

              How to Disconnect a Philips Respironics V60               
  • Unplug the ventilator

Unplug the ventilator

Step 4

  • Disconnect the internal battery from the power harness connector.

Disconnect the internal battery from the power harness connector.

Step 5

              Top Cover               
  • Turn the captive D-ring fastener on the bottom of the panel counter-clockwise one-quarter turn and release.
  • Then remove the right side panel.

Turn the captive D-ring fastener on the bottom of the panel counter-clockwise one-quarter turn and release.

Then remove the right side panel.

Step 6

  • Remove the 4 screws using a screwdriver.
  • Remove the top cover.

Remove the 4 screws using a screwdriver.

Remove the top cover.

To reassemble your device, follow these instructions in reverse order.

Cancel: I did not complete this guide.