Introduction

Proper Use:

  • The exhalation block is for a single patient only and must be replaced for each new patient.It can be removed from the ventilator and cleaned for use on the same patient by washing with soap and water.If cleaning the exhalation block, ensure that it is completely dry before reinserting it back into the ventilator. This ensures that no damage is done to the ventilator electronics.The exhalation block can either be left to air-dry, or blown dry with compressed air.

The repair information is also available on pages 144-145 (Section 5, Pages 14-15) of the Service Manual that is linked in the documents section.

Step 1

              Exhalation Block               
  • Locate the exhalation block on the front of the ventilator.

Locate the exhalation block on the front of the ventilator.

1024

Step 2

  • Tilt the machine up and unscrew the knob that holds the exhalation block in place.
  • This knob can be untwisted by hand, no tools required.

Tilt the machine up and unscrew the knob that holds the exhalation block in place.

This knob can be untwisted by hand, no tools required.

Step 3

  • Pinch the “From Patient” fitting and slide the exhalation block away from the ventilator.

Pinch the “From Patient” fitting and slide the exhalation block away from the ventilator.

Step 4

  • After replacing or cleaning the block, perform the exhalation flow sensor calibration.
  • See 6.7.3.3 in the Service Manual for instructions on calibration.

After replacing or cleaning the block, perform the exhalation flow sensor calibration.

See 6.7.3.3 in the Service Manual for instructions on calibration.

To reassemble your device, follow these instructions in reverse order.

Cancel: I did not complete this guide.

                                                                                      2 other people completed this guide.                                             

Author

                                      with 2 other contributors 

                    Kyle Hammer                     

Member since: 04/07/2020

292 Reputation

                                      2 Guides authored                  



                       Badges:
                       7







                                                        +4 more badges                           

Team

                       Cal Poly, Team S11-G20, Paton Spring 2020                        

                                                  Member of Cal Poly, Team S11-G20, Paton Spring 2020 



                    CPSU-PATON-S20S11G20                     


                                            2 Members                     


                                            3 Guides authored