Introduction
The BPS battery pack can be found in one of two places. If the ventilator stands on a pole connected to four wheels, it does not have a compressor, and the battery is located directly above the wheels. If the ventilator does not stand on a single pole, the battery is located in the compressor mount cart. This guide covers the replacement of the BPS battery in ventilators with compressors.
For more information on replacing the BPS battery, please refer to page 8-114 (section 8.18.3) in the service manual.
Step 1
BPS Battery
- Before beginning, ensure that the battery is drained and the ventilator is not connected to a power source.
- Locate the breath delivery unit (BDU) shelf on the ventilator.
- Locate the battery indicator label on the shelf to determine whether the BPS battery holds a one-hour or four-hour charge - the default for this guide is a four-hour battery.
Before beginning, ensure that the battery is drained and the ventilator is not connected to a power source.
Locate the breath delivery unit (BDU) shelf on the ventilator.
Locate the battery indicator label on the shelf to determine whether the BPS battery holds a one-hour or four-hour charge - the default for this guide is a four-hour battery.
1024
Step 2
- Grasp the sides of the front bezel of the BPU and gently spread them away from the ventilator.
- Pull the bezel outward to remove it from the cart.
- Use a Phillips screwdriver to remove the two screws that connect the BPS to the BDU shelf at the back of the cart.
Grasp the sides of the front bezel of the BPU and gently spread them away from the ventilator.
Pull the bezel outward to remove it from the cart.
Use a Phillips screwdriver to remove the two screws that connect the BPS to the BDU shelf at the back of the cart.
Step 3
- Using a Phillips screwdriver, remove the eleven screws securing the BPS top and front metal covers.
- If the BPS contains a one-hour battery, only the four screws holding the one-hour battery shield (four on the top cover) need to be removed.
- Pull the covers away from the BPS.
Using a Phillips screwdriver, remove the eleven screws securing the BPS top and front metal covers.
If the BPS contains a one-hour battery, only the four screws holding the one-hour battery shield (four on the top cover) need to be removed.
Pull the covers away from the BPS.
Step 4
- Locate the battery indicator bracket.
- Cut the zip tie connecting the battery harness (bundle of wires coming from the battery) to the battery indicator bracket.
- If the ventilator contains a one-hour battery, simply disconnect the battery harness from the battery cable and lift the battery out of the BPS.
Locate the battery indicator bracket.
Cut the zip tie connecting the battery harness (bundle of wires coming from the battery) to the battery indicator bracket.
If the ventilator contains a one-hour battery, simply disconnect the battery harness from the battery cable and lift the battery out of the BPS.
Step 5
- Use a Phillips screwdriver to remove the two screws securing the battery retaining bracket to the rear of the BPS.
Use a Phillips screwdriver to remove the two screws securing the battery retaining bracket to the rear of the BPS.
Step 6
- Disconnect the battery connector from J2 on the circuit board next to the battery indicator bracket.
Disconnect the battery connector from J2 on the circuit board next to the battery indicator bracket.
Step 7
- Using a Phillips screwdriver, loosen the two side screws and the two base screws holding the BPS front bracket.
- Carefully lift the battery out of the BPS.
Using a Phillips screwdriver, loosen the two side screws and the two base screws holding the BPS front bracket.
Carefully lift the battery out of the BPS.
To reassemble your device, follow these instructions in reverse order.
Cancel: I did not complete this guide.
2 other people completed this guide.
Author
with 1 other contributor
Olivia May
Member since: 01/23/2020
510 Reputation
1 Guide authored
Badges:
8
+5 more badges
Team
Cal Poly, Team S7-G4, Paton Spring 2020
Member of Cal Poly, Team S7-G4, Paton Spring 2020
CPSU-PATON-S20S7G4
3 Members
3 Guides authored