Introduction
What you need
Step 1
Children/Adult Expiratory Filter
- Do not begin process while patient is connected. Alarm will sound
- If necessary, remove the expiratory limb (item 4) of the patient circuit from the exhalation filter.
Do not begin process while patient is connected. Alarm will sound
If necessary, remove the expiratory limb (item 4) of the patient circuit from the exhalation filter.
1024
Step 2
- Raise the exhalation filter latch to unlock (item 3).
Raise the exhalation filter latch to unlock (item 3).
Step 3
- Open the exhalation filter door (item 4).
Open the exhalation filter door (item 4).
Step 4
- Remove the existing filter. Dispose of properly as per institution’s guidelines on disposing hazardous waste.
Remove the existing filter. Dispose of properly as per institution’s guidelines on disposing hazardous waste.
Step 5
- Ensure filter is properly attached to ventilator
- Only use filters designed for Puritan Bennett 980 Series Ventilator
- Insert the new filter by sliding the filter along the tracks in the door. Ensure the from patient port (item 6) aligns with the cutout in the door and points away from the ventilator.
- Note: The condensate vial (item 5) is removable for discarding accumulated liquid. Turn this vial clockwise to remove. Turn this vial counterclockwise to install.
Ensure filter is properly attached to ventilator
Only use filters designed for Puritan Bennett 980 Series Ventilator
Insert the new filter by sliding the filter along the tracks in the door. Ensure the from patient port (item 6) aligns with the cutout in the door and points away from the ventilator.
Note: The condensate vial (item 5) is removable for discarding accumulated liquid. Turn this vial clockwise to remove. Turn this vial counterclockwise to install.
Step 6
- Close the exhalation filter door (item 4).
Close the exhalation filter door (item 4).
Step 7
- Lower the exhalation filter latch to secure the filter (item 3).
Lower the exhalation filter latch to secure the filter (item 3).
Cancel: I did not complete this guide.
One other person completed this guide.
Author
with 1 other contributor
Silvia Calinov
Member since: 04/08/2020
175 Reputation
1 Guide authored
Badges:
3
Team
Cal Poly, Team S11-G14, Paton Spring 2020
Member of Cal Poly, Team S11-G14, Paton Spring 2020
CPSU-PATON-S20S11G14
3 Members
4 Guides authored