Difficulty

Easy

Steps

7

Time Required

                          3 - 5 minutes            

Sections

1

  • Children/Adult Expiratory Filter
  • 7 steps

Flags

0

  • BackPuritan Bennett 980

  • Full Screen

  • Options

  • History

  • Save to Favorites

  • Download PDF

  • Edit

  • Translate

  • Get Shareable Link

  • Embed This Guide

  • Notify Me of Changes

  • Stop Notifications

Introduction

What you need

Step 1

              Children/Adult Expiratory Filter               
  • Do not begin process while patient is connected. Alarm will sound
  • If necessary, remove the expiratory limb (item 4) of the patient circuit from the exhalation filter.

Do not begin process while patient is connected. Alarm will sound

If necessary, remove the expiratory limb (item 4) of the patient circuit from the exhalation filter.

1024

Step 2

  • Raise the exhalation filter latch to unlock (item 3).

Raise the exhalation filter latch to unlock (item 3).

Step 3

  • Open the exhalation filter door (item 4).

Open the exhalation filter door (item 4).

Step 4

  • Remove the existing filter. Dispose of properly as per institution’s guidelines on disposing hazardous waste.

Remove the existing filter. Dispose of properly as per institution’s guidelines on disposing hazardous waste.

Step 5

  • Ensure filter is properly attached to ventilator
  • Only use filters designed for Puritan Bennett 980 Series Ventilator
  • Insert the new filter by sliding the filter along the tracks in the door. Ensure the from patient port (item 6) aligns with the cutout in the door and points away from the ventilator.
  • Note: The condensate vial (item 5) is removable for discarding accumulated liquid. Turn this vial clockwise to remove. Turn this vial counterclockwise to install.

Ensure filter is properly attached to ventilator

Only use filters designed for Puritan Bennett 980 Series Ventilator

Insert the new filter by sliding the filter along the tracks in the door. Ensure the from patient port (item 6) aligns with the cutout in the door and points away from the ventilator.

Note: The condensate vial (item 5) is removable for discarding accumulated liquid. Turn this vial clockwise to remove. Turn this vial counterclockwise to install.

Step 6

  • Close the exhalation filter door (item 4).

Close the exhalation filter door (item 4).

Step 7

  • Lower the exhalation filter latch to secure the filter (item 3).

Lower the exhalation filter latch to secure the filter (item 3).

Cancel: I did not complete this guide.

                                                                                      One other person completed this guide.                                             

Author

                                      with 1 other contributor 

                    Silvia Calinov                     

Member since: 04/08/2020

175 Reputation

                                      1 Guide authored                  



                       Badges:
                       3

Team

                       Cal Poly, Team S11-G14, Paton Spring 2020                        

                                                  Member of Cal Poly, Team S11-G14, Paton Spring 2020 



                    CPSU-PATON-S20S11G14                     


                                            3 Members                     


                                            4 Guides authored