Difficulty

Moderate

Steps

17

Time Required

                          50 minutes            

Sections

1

  • Pump Motor
  • 17 steps

Flags

0

  • BackResMed AirSense 10

  • Full Screen

  • Options

  • History

  • Save to Favorites

  • Download PDF

  • Edit

  • Translate

  • Get Shareable Link

  • Embed This Guide

  • Notify Me of Changes

  • Stop Notifications

Introduction

What you need

Step 1

              Pump Motor               
  • Position the ResMed Airsense 10 with the water reservoir facing towards you.
  • Grip both the top and bottom of the water reservoir and pull it straight out and away from the device.

Position the ResMed Airsense 10 with the water reservoir facing towards you.

Grip both the top and bottom of the water reservoir and pull it straight out and away from the device.

1024

Step 2

  • Position the ResMed Airsense 10 with the screen facing you to remove the front cover plate.
  • Place your finger under the indent beneath the home button and pry the front cover plate off.
  • This should only require medium pressure.

Position the ResMed Airsense 10 with the screen facing you to remove the front cover plate.

Place your finger under the indent beneath the home button and pry the front cover plate off.

This should only require medium pressure.

Step 3

  • Remove the two 14mm Torx T10 screws that were exposed by removing the front cover plate.

Remove the two 14mm Torx T10 screws that were exposed by removing the front cover plate.

Step 4

  • Rotate the device to face the bottom of the ResMed Airsense 10.
  • Remove the four 10mm Torx T10 screws, then place the screws and the plastic heater plate cover to the side.
  • Remove the remaining two 14mm Torx T10 screws.

Rotate the device to face the bottom of the ResMed Airsense 10.

Remove the four 10mm Torx T10 screws, then place the screws and the plastic heater plate cover to the side.

Remove the remaining two 14mm Torx T10 screws.

Step 5

  • Once you have removed the cover of the heating plate, disconnect the orange wire from the fuse.
  • Carefully pull the orange wire through the silicone gasket.
  • Fully remove the heater plate from the ResMed Airsense 10.

Once you have removed the cover of the heating plate, disconnect the orange wire from the fuse.

Carefully pull the orange wire through the silicone gasket.

Fully remove the heater plate from the ResMed Airsense 10.

Step 6

  • Remove the remaining two 14mm Torx T10 screws holding the frame together.

Remove the remaining two 14mm Torx T10 screws holding the frame together.

Step 7

  • Orient the ResMed AirSense 10 to face the rear of the machine. The air outlet should be facing towards you.
  • Use the plastic spudger to pry the casing of the ResMed AirSense 10 off of the unit.
  • Fully remove the casing from the ResMed frame.

Orient the ResMed AirSense 10 to face the rear of the machine. The air outlet should be facing towards you.

Use the plastic spudger to pry the casing of the ResMed AirSense 10 off of the unit.

Fully remove the casing from the ResMed frame.

Step 8

  • Position the ResMed AirSense 10 with the screen facing towards you.
  • Carefully remove the 4 pin wire connector located to the left of the screen.
  • Fully remove left panel of the ResMed.

Position the ResMed AirSense 10 with the screen facing towards you.

Carefully remove the 4 pin wire connector located to the left of the screen.

Fully remove left panel of the ResMed.

Step 9

  • Position the ResMed AirSense 10 with the air outlet facing you.
  • Remove the 14mm Torx T10 screw that is located between the air outlet and the front screen.
  • Remove the wires from the pump casing’s hook.
  • Carefully start to remove the pump casing from the Resmed AirSense 10 until the wire connector is exposed.

Position the ResMed AirSense 10 with the air outlet facing you.

Remove the 14mm Torx T10 screw that is located between the air outlet and the front screen.

Remove the wires from the pump casing’s hook.

Carefully start to remove the pump casing from the Resmed AirSense 10 until the wire connector is exposed.

Step 10

  • Unclamp the connector for the wires that connect the pump casing to the motherboard to fully disconnect the pump.

Unclamp the connector for the wires that connect the pump casing to the motherboard to fully disconnect the pump.

Step 11

  • Orient the pump casing with the wires facing towards you.
  • Unhook the wires from the outer most clip.
  • Then remove the plastic retainer from the pump casing.

Orient the pump casing with the wires facing towards you.

Unhook the wires from the outer most clip.

Then remove the plastic retainer from the pump casing.

Step 12

  • Remove the blue silicone buffer to right corner.
  • Remove the wires from the remaining hooks.

Remove the blue silicone buffer to right corner.

Remove the wires from the remaining hooks.

Step 13

  • Orient the pump casing to face the outlet port.
  • Push the outlet nipple into the pump housing.

Orient the pump casing to face the outlet port.

Push the outlet nipple into the pump housing.

Step 14

  • Orient the pump casing to face the three T10 torx screws.
  • Remove the three 14mm Torx T10 screws.
  • Lift off the top of the pump casing.

Orient the pump casing to face the three T10 torx screws.

Remove the three 14mm Torx T10 screws.

Lift off the top of the pump casing.

Step 15

  • Remove the pump assembly from the pump casing and face the pump assembly with the mounting plate facing you.
  • With the pump assembly out of the pump, there is a mounting plate that needs to be removed to access the pump.
  • There are 5 rubber ridges connecting the mounting plate to the rubber sock encasing the pump.
  • Push the ridges through the holes in the mounting plate.

Remove the pump assembly from the pump casing and face the pump assembly with the mounting plate facing you.

With the pump assembly out of the pump, there is a mounting plate that needs to be removed to access the pump.

There are 5 rubber ridges connecting the mounting plate to the rubber sock encasing the pump.

Push the ridges through the holes in the mounting plate.

Step 16

  • Separate the mounting plate from the rubber sock.

Separate the mounting plate from the rubber sock.

Step 17

  • Pull off the rubber sock to expose the pump.
  • Do not damage the wires while removing the rubber sock.
  • The exposed pump can now be cleaned and inspected for damage.

Pull off the rubber sock to expose the pump.

Do not damage the wires while removing the rubber sock.

The exposed pump can now be cleaned and inspected for damage.

To reassemble your device, follow these instructions in reverse order.

Cancel: I did not complete this guide.

                                                                                      14 other people completed this guide.                                             

Author

                                      with 4 other contributors 

                    Jacey Cobb                     

Member since: 09/16/2020

977 Reputation

                                      3 Guides authored                  



                       Badges:
                       11







                                                        +8 more badges                           

Team

                       The Citadel Military College of South Carolina, Team S1-G6, Eggleston Fall 2020                        

                                                  Member of The Citadel Military College of South Carolina, Team S1-G6, Eggleston Fall 2020 



                    CMCSC-EGGLESTON-F20S1G6                     


                                            3 Members                     


                                            6 Guides authored                     

glove2006 - Feb 12, 2021

Reply

Thank you @jaceyt ! This was spot on.

Shayne Braid - Feb 23, 2021

Reply

Where can i buy a replacement pump?

Travis Verner - Aug 27, 2022

Alibaba sells the pump. Everyone else is OOS.

christecnico - Apr 7, 2021

Reply

Where can i buy replacements?

jglazer - Apr 20, 2021

Reply

Step 1, find a replacement motor???