- Remove the water tub by gently pulling it out.
Remove the water tub by gently pulling it out.
1024
Step 2
- Pry off the face plate by using your finger or an opening tool.
- Use a moderate amount of force when prying open.
- Remove the face plate from the top cover.
- There will be a clicking sound on either side of the divot when assembling and disassembling.
Pry off the face plate by using your finger or an opening tool.
Use a moderate amount of force when prying open.
Remove the face plate from the top cover.
There will be a clicking sound on either side of the divot when assembling and disassembling.
Step 3
- Unscrew and remove the two 11.8 mm TR10 Torx screws that were revealed from the face plate being removed.
Unscrew and remove the two 11.8 mm TR10 Torx screws that were revealed from the face plate being removed.
Step 4
- Unscrew and remove the two 11.8 mm TR10 Torx screws from the bottom of the device.
Unscrew and remove the two 11.8 mm TR10 Torx screws from the bottom of the device.
Step 5
- Remove the top cover from the rest of the device.
Remove the top cover from the rest of the device.
To reassemble your device, follow these instructions in reverse order.
Cancel: I did not complete this guide.
Author
with 3 other contributors
Mitchell Novak
Member since: 01/16/2020
760 Reputation
5 Guides authored
Badges:
11
+8 more badges
Team
Embry-Riddle Aeronautical University, Team 1-1, Rauch Summer 2022
Member of Embry-Riddle Aeronautical University, Team 1-1, Rauch Summer 2022
ERAU-RAUCH-SU22S1G1
3 Members
5 Guides authored