Introduction

Preparations:

• Make sure that the Servo-i works properly before performing any maintenance.

• Set the On/Off switch on the User Interface to Off.

• Disconnect the mains power cable.

• Disconnect the gas supplies (wall and/or cylinder).

• Remove patient tubing.

• If fitted, remove bacteria filter from the expiratory inlet

General:

• A ‘Preventive maintenance’ must be performed at least once every year as long as the unit is not used more than normal. Normal operation is estimated to correspond to approx. 5,000 hours of operation.

• The Battery modules shall be replaced every three years.

• The internal Lithium batteries on PC 1771 and

PC 1772 shall be replaced every five years.

• After or in combination with performing the ‘Preventive maintenance’ described in this chapter, a ‘Routine cleaning’ and an ‘Extended cleaning of Inspiratory channel’ must be performed. For cleaning procedures, refer to the ‘Servo-i Ventilator System – User’s manual’.

• In some parts of the Preventive maintenance, as described in this chapter, access to the Field Service System (FSS) is required.

• Some optional equipment used with the Servo-i, but not covered by this Service Manual, may also demand maintenance actions. Refer to the optional equipments documentation.

Example on such optional equipment are:

– Humidifier

– Compressor Mini.

Equipment:

• Standard service tools.

• Barometer (or information about the actual barometric pressure).

• Access to the Field Service System (FSS ). Recommended but not required.

• Preventive maintenance-kit containing all parts

needed during the maintenance.

(Service Manual 7 - 3)

What you need

Step 1

              Filter for Gas Modules               
  • When performing this maintenance, a ‘Maintenance kit, 5,000 hours’ should be used
  • Remove the inspiratory section cover
  • Loosen the screw (1)
  • Lift off the gas module bracket (2)
  • Service Manual 4 - 14

When performing this maintenance, a ‘Maintenance kit, 5,000 hours’ should be used

Remove the inspiratory section cover

Loosen the screw (1)

Lift off the gas module bracket (2)

Service Manual 4 - 14

1024

Step 2

  • Pull out and lift off both gas modules.
  • Service Manual 4 - 14

Pull out and lift off both gas modules.

Step 3

  • Note 1: When assembling, make sure that the connector muff properly seals around the inspiratory pipe (1). The connector muff must not be pushed too far onto the nozzle units (2)
  • Note 2: The gas modules used in Servo-i are factory adjusted for this purpose. When replacing gas modules, make sure to use only Servo-i gas modules.Similar gas modules intended for the SV 300/300A or KION / KION-i Systems must not be used.
  • Service Manual 4 - 14

Note 1: When assembling, make sure that the connector muff properly seals around the inspiratory pipe (1). The connector muff must not be pushed too far onto the nozzle units (2)

Note 2: The gas modules used in Servo-i are factory adjusted for this purpose. When replacing gas modules, make sure to use only Servo-i gas modules.Similar gas modules intended for the SV 300/300A or KION / KION-i Systems must not be used.

Step 4

  • Replace filters in the gas modules. When replacing filter, move the rubber seal from the old to the new filter.
  • Service Manual 7 - 3

Replace filters in the gas modules. When replacing filter, move the rubber seal from the old to the new filter.

Service Manual 7 - 3

Step 5

              Bacteria Filter for O2 Cell               
  • Twist off white filter cap.
  • Service Manual 7 - 4

Twist off white filter cap.

Service Manual 7 - 4

Step 6

  • Use the Maintenance kit, 5,000 hours to replace the bacteria filter for the O2 cell.
  • Service Manual 7 - 4

Use the Maintenance kit, 5,000 hours to replace the bacteria filter for the O2 cell.

To reassemble your device, follow these instructions in reverse order.

Cancel: I did not complete this guide.

Author

                                      with 3 other contributors 

                    rachel davis                     

Member since: 04/19/2018

107 Reputation

                                      1 Guide authored                  



                       Badges:
                       8







                                                        +5 more badges                           

Team

                       Cal Poly, Team S1-G2, Zimmerman Spring 2020                        

                                                  Member of Cal Poly, Team S1-G2, Zimmerman Spring 2020 



                    CPSU-ZIMMERMAN-S20S1G2                     


                                            5 Members                     


                                            5 Guides authored